High speed centrifuge use code

1. When the centrifuge is in the pre-cooling state, the centrifuge lid must be closed. After the centrifugation is finished, the rotor should be placed on the test bench, and the residual water in the chamber should be dried. The centrifuge lid is open.

2. Rotating head When pre-cooling, the rotating head cover can be placed on the platform of the centrifuge, or placed on the test bench. Do not screw it on the rotating head, because once it is accidentally activated, the rotating head cover Will fly out, causing an accident!

3. After the screw cap is tightened, be sure to touch the gap between the rotor and the cover with your fingers. If there is a gap, unscrew and retighten until the gap is confirmed before the centrifuge can be started.

4. During the centrifugation process, the operator must not leave the centrifuge room. Once the abnormal situation occurs, the operator cannot turn off the power (POWER), press STOP. Fill in the centrifuge usage record before pre-cooling.

5. Do not use fake and inferior centrifugation to make the tube, and do not use aging, deformed, cracked centrifuge tubes.

6.
Can leave after the last routine safety check using a centrifuge on holidays and evenings .

7. In the event of machine failure during the use of the instrument, contact the manufacturer in time for damage to the component.