Attention when using carbon dioxide incubator

1. The carbon dioxide incubator cannot be turned on before the water is injected, otherwise the heating element will be damaged.

2. After running the CO2 incubator for several months, the water in the tank may be reduced due to volatilization. When the low water level indicator (WLow12) is on, it should be supplemented with water. First open the overflow pipe, use the funnel to connect the rubber tube to add water from the water injection hole to extinguish the low water level indicator, and then add the supplementary water (CP-ST200A plus water 1800ml, CP-ST100A plus water 1200ml), and then block the overflow hole.

3. The carbon dioxide incubator can be used in a high-precision constant temperature incubator. At this time, the CO2 control system must be turned off.

4. Since the CO2 sensor is calibrated under saturated humidity, the humidifying disc must always be filled with sterilized water.

5. When the display temperature exceeds the set temperature by 1 °C, the over temperature alarm indicator (OverTemp7) lights up and a sharp alarm sounds. At this time, the power should be turned off for 30 minutes; if the power (temperature control) switch (Power20) is turned on again If it is over temperature, the power should be turned off and reported to the maintenance personnel.

6. When the cylinder pressure is lower than 0.2MPa, the cylinder should be replaced.

7. Minimize the time to open the glass door.

8. If the carbon dioxide incubator is not used for a long time, the moisture in the working room must be removed before closing. Open the glass door and ventilate for 24 hours before closing.

9. When cleaning the incubator studio, do not hit the sensor and the stirring motor wind wheel and other components.

10, disassembly and assembly of the indoor support bracket, must use a random special wrench, not excessive force 11. Before transporting the carbon dioxide incubator, the water in the tank must be removed. When draining, put the rubber tube tightly on the water outlet hole so that the nozzle is lower than the instrument, gently suck a bit, and put down the water pipe, the water is siphone out.

12. Before carrying the carbon dioxide incubator, take out the shelf and humidifying tray in the working chamber to prevent the glass door from colliding.

13. When carrying the carbon dioxide incubator, do not invert it, and do not lift the door to avoid deformation of the door.

Fruit Juice Powder

China Fruit Juice Powder,Cranberry Fruit Powder manufacturers, welcome Tart Cherry Powder,Passion Fruit Powder purchasers from worldwide to visit our site.

Our healthway supply Medical Intermediate Herbal Extracts Natural Plant Herbal Extract Super Food Ingredients Fruit Powder Fruit Juice Powder Freeze-dried Powder Vegetable Powder Food Additives Cosmestic Ingredienrs
Icariin NMN powder Cissus Irish sea moss Quercetin Barley grasss, We had passed FDA certificate and ISO 9001 .

Fruit Juice Powder,Cranberry Fruit Powder,Tart Cherry Powder,Passion Fruit Powder

Xi'an Healthway Biotech Co.,Ltd , https://www.xahealthway.com